Instructions

RIGOL Chapter 2 Command System
2-76 DSG800 Programming Guide
[:SOURce]:SWEep:SWEep:TRIGger:TYPE
Syntax [:SOURce]:SWEep:SWEep:TRIGger:TYPE AUTO|KEY|BUS|EXT
[:SOURce]:SWEep:SWEep:TRIGger:TYPE?
Description Set the trigger mode of the sweep period.
Query the trigger mode of the sweep period.
Parameter
Name Type Range Default
AUTO|KEY|BUS|EXT Discrete AUTO|KEY|BUS|EXT AUTO
Explanation AUTO: select "Auto" trigger mode. If the sweep mode is set to "Cont", the
instrument will start sweeping once a sweep manner is selected. If the sweep
mode is set to "Single", you need to send the
[:SOURce]:SWEep:EXECute
command to meet the single sweep condition; after that, the instrument will
start a sweep and then stops.
KEY: select "Key" trigger mode. If the sweep mode is set to "Cont", the
instrument starts a sweep each time Key Trig is pressed; if the sweep mode is
set to "Single", you need to send the
[:SOURce]:SWEep:EXECute command to
meet the single sweep condition and after that, the instrument starts a sweep
and then stops each time Key Trig is pressed.
BUS: select "Bus" trigger mode. If the sweep mode is set to "Cont", the
instrument starts a sweep each time the
*TRG
or
:TRIGger[:SWEep][:IMMediate] command is sent; if the sweep mode is set to
"Single", you need to send the
[:SOURce]:SWEep:EXECute command to meet
the single sweep condition and after that, the instrument starts a sweep and
then stops each time the
*TRG or :TRIGger[:SWEep][:IMMediate] command is
sent.
EXT: select "Ext" trigger mode. The RF signal generator receives the trigger
signal input from the [TRIGGER IN] connector at the rear panel. If the sweep
mode is set to "Cont", the instrument starts a sweep each time a TTL pulse
signal with the specified polarity is received. If the sweep mode is set to
"Single", you need to send the
[:SOURce]:SWEep:EXECute command to meet
the single sweep condition; after that, the instrument starts a sweep and then
stops each time a TTL pulse signal with the specified polarity is received.
Note: The above explanations are only valid when the trigger mode of each
sweep point within the sweep period is met.
When executing the sweep operation, the priority of the required conditions is:
single sweep trigger mode point trigger mode.
Return Format The query returns the trigger mode of the sweep. For example, AUTO.
Example
:SWE:SWE:TRIG:TYPE AUTO
:SWE:SWE:TRIG:TYPE?
Related
Commands
[:SOURce]:SWEep:EXECute
[:SOURce]:SWEep:MODE
[:SOURce]:SWEep:POINt:TRIGger:TYPE
*TRG
:TRIGger[:SWEep][:IMMediate]