Instructions
RIGOL Chapter 2 Front Panel Operations
2-10 DSG800 User Guide
F/L
T
Start Frequency
/Start Level
Start Frequency
/Start Level
Ramp
Triangle
Stop Frequency
/Stop Level
Stop Frequency
/Stop Level
Ramp: the sweep period always starts from the start frequency or start level
to the stop frequency or stop level and the sweep sequence is similar to a
ramp waveform.
Triangle: the sweep period always starts from the start frequency or start
level to the stop frequency or stop level and then returns back to the start
frequency or start level. The sweep sequence is similar to a triangle
waveform.
Sweep Mode
Press SWEEP Mode to select "Cont" or "Single" sweep and the default is "Cont".
Cont: when continuous sweep is selected, the continuous sweep label is
displayed in the frequency or amplitude area in the user interface. When the
trigger condition is met, the instrument sweeps continuously according to the
current setting.
Single: when single sweep is selected, the single sweep label is displayed in the
frequency or amplitude area in the user interface. When the trigger condition is
met, the instrument performs a single sweep according to the current setting
and then stops.
Single Sweep
If the current sweep mode is "Cont", press Single to switch the sweep mode to
"Single" and the instrument will perform a single sweep if the trigger condition is