Data Sheet

Specications
All the specications are guaranteed except parameters marked with “Typical” and the oscilloscope needs to operate for more than 30 minutes
under the specied operation temperature.
Sample
Sample Mode Real-time sample
Real-time Sample Rate
Analog channel:
1 GSa/s (single-channel),
500 MSa/s (dual-channel), 250 MSa/s (three/four-channel)
Digital channel:
1 GSa/s (8-channel), 500 MSa/s (16-channel)
Peak Detect
Analog channel: 4 ns
Digital channel: 4 ns
Averaging After all the channels nish N samples at the same time, N can be 2, 4, 8, 16, 32, 64, 128, 256, 512 or 1024.
High Resolution 12 bit (max.)
Interpolation Sin(x)/x (optional)
Minimum Detect Pulse Width Digital channel: 10 ns
Memory Depth
Analog channel:
standard 12 Mpts (single-channel), 6 Mpts (dual-channel), 3 Mpts (three/four-channel); optional 24 Mpts
(single-channel), 12 Mpts (dual-channel), 6 Mpts (three/four-channel)
Digital channel:
standard 12 Mpts (8-channel), 6 Mpts (16-channel); optional 24 Mpts (8-channel), 12 Mpts (16-channel)
Input
Number of Channels
MSO1XX4Z/1XX4Z-S: 4 analog channels, 3 analog channels+8 digital channels,
2 analog channels+16 digital channels
DS1XX4Z Plus/1XX4Z-S Plus: 4 analog channels, MSO upgradable
DS1054Z: 4 analog channels
Input Coupling DC, AC or GND
Input Impedance
Analog channel: (1 MΩ±1%) || (15 pF±3 pF)
Digital channel: (100 kΩ±1%) || 8 pF±3 pF)
Probe Attenuation
Coefcient
Analog channe: 0.01X to 1000X, in 1-2-5 step
Maximum Input Voltage
(1 MΩ)
Analog channel:
CAT I 300 Vrms, CAT II 100 Vrms, transient overvoltage 1000 Vpk
With RP2200 10:1 probe: CAT II 300 Vrms
Digital channel:
CAT I 40 Vrms, transient overvoltage 800 Vpk
Horizontal
Timebase Scale 5 ns/div to 50 s/div
Maximum Record Length 24 Mpts (optional)
Timebase Accuracy
[1]
≤ ± 25 ppm
Clock Drift ≤ ± 5 ppm/year
Maximum Delay Range
Negative delay: 1/2 (Memory Depth/Sample Rate)
Positive delay: 1 s to 500 s
Timebase Mode YT, XY, Roll
Number of X-Ys 1
Waveform Capture Rate
[2]
30,000 wfms/s (dots display)
Zero Offset ±0.5div*minimum time base scale
Vertical
Bandwidth (-3dB)
MSO1104Z/1104Z-S and DS1104Z Plus/1104Z-S Plus: DC to 100 MHz
MSO1074Z/1074Z-S and DS1074Z Plus/1074Z-S Plus: DC to 70 MHz
DS1054Z: DC to 50 MHz