Data Sheet

5 To Trigger the Oscilloscope RIGOL
MSO4000D/S4000 Users Guide
5-23
pattern, the former edge item defined will be replaced by X.
Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,
normal or single. The corresponding status light of the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger coupling, trigger holdoff and
noise rejection) under this trigger type.
Trigger Level:
Use TRIGGER
LEVEL to modify the level. For details, please refer to the
description of
Trigger Levelon page 5-9. Note that the trigger level of each
channel needs to be set respectively.