Data Sheet

6 To Make Measurements RIGOL
MSO4000D/S4000 Users Guide
6-9
Press MATH Operate to select “Logic”:
Press Log.Formula to select the desired operation expression and the default is
AND.
Press Source A and Source B to select the desired channels. The channels
available are CH1, CH2, CH3 and CH4 or D0-D15.
Press and use to adjust the vertical position of the operation results.
Press
and use to adjust the vertical scale of the operation results.
Press Invert to turn the inverted display of the operation results on or off.
Press Threshold A and use to set the threshold of source A in logic
operation.
Press Threshold B and use to set the threshold of source B in logic
operation.
HORIZONTAL
POSITION and HORIZONTAL SCALE can also be
used to adjust the horizontal position and scale of the operation results.