Data Sheet

RIGOL 6 To Make Measurements
MSO4000/DS4000 Users Guide
6-20
Measurement Setting
1. Source Selection
Press Measure Source to select the desired channel for measurement (CH1
to CH4, MATH or D0-D15). The color of the parameter icons under MENU at the
left of the screen will change with the source selected.
2. Measurement Range
Press Measure Meas.Range Meas.Range to select “Screen Regionor
Cursor Regionfor measurement. When Cursor Regionis selected, two cursor
lines appear on the screen. At this point, press Cursor A and Cursor B and use
to adjust the position of the two cursor lines respectively so as to determine
the measurement range. Or press CursorAB and use to adjust the
positions of the cursor A and B at the same time. Note that you can press
continuously to switch between the cursors.
3. Delayed Measurement Setting
Specify the source A and source B in the measurement items “Delay AB
and “Delay A
B ”.
Press Measure Meas.Setting Type Delayand then press Source A
and Source B to set the two channel sources (CH1 to CH4, MATH or D0-D15) of
delayed measurement respectively.