Data Sheet

6 To Make Measurements RIGOL
MSO4000D/S4000 Users Guide
6-35
X-Y Mode
The X-Y mode is only avalibale when the main time base mode is “X-Y”. In this mode,
you can adjust the cursor positions to measure the X and Y values at the crossing
points of the two pair of cursors.
Press MENU in the horizontal control area (HORIZONTAL) at the front panel and
press Time Base to select “X-Y. Then, press Cursor Mode “X-Y” to enable
the “X-Y” cursor measurement function. The measurement results are displayed at
the upper-left cornor of the scree, as shown in the figure below.
CurAX: the X value at cursor A.
CurBX: the X value at cursor B.
CurAY: the Y value at cursor A.
CurBY: the Y value at cursor B.
You can adjust the cursor positions in the following way:
1. Select cursor region
Press Region to select “XY1” or “XY2”.
2. Select cursor type and adjust cursor positions
Press Type to select “X”. Press CursorA and use to adjust the
horizontal position of cursor A. Press CursorB and use to adjust the
horizontal positon of cursor B. Press CursorAB and use to adjust the
horizontal positions of cursor A and cursor B at the same time. During the
adjustment, the measurement results will change accordingly and the
adjustable range is limited within the screen.
Press Type to select “Y”. Press CursorA and use to adjust the vertical
position of cursor A. Press CursorB and use to adjust the vertical
positon of cursor B. Press CursorAB and use to adjust the vertical
positions of cursor A and cursor B at the same time. During the adjustment,