Data Sheet

RIGOL 8 Protocol Decoding
MSO4000/DS4000 Users Guide
8-16
CAN Decoding (Option)
Press Decode1 Decode and select “CAN” to open the CAN decoding function
menu.
1. Source
Press Source and select any channel (CH1-CH4 or D0-D15) as the source
channel.
2. Signal Type
Press Signal Type to select the desired signal type.
CAN_H: the actual CAN_H bus signal.
CAN_L: the actual CAN_L bus signal.
Differential: the CAN differential bus signals connected to an analog
channel using a differential probe. The positive lead of the probe connects
CAN_H and the negative lead connects CAN_L.
3. Baud
Press Baud to select a baud rate (100 kb/s, 125 kb/s, 250 kb/s, 400 kb/s, 500
kb/s, 800 kb/s, 1 Mb/s or User) that matches the CAN bus signal. When Useris
selected, press Setup and use to enter the desired rate, the range is from
10 kb/s to 1 Mb/s.
4. Sample Point
The Sample point is a point within a bits time. The oscilloscope samples the bit
level at this point. “Sample pointis represented by the percentage of the time
from the start of the bits time to the sample point timein the bits time.
Press Sample Point and use to adjust this parameter with a step of 1%.
The range is from 5% to 95%.
1bit
60%
70%
80%