4. Create Mask
The mask is a template used to perform waveform analysis. Press Create Mask to
immediately apply the mask (X Mask and Y Mask) currently created.
Test and Ouput
Before the test, you can use the following method to set the output mode of the test
results.
Press Msg Display to selectONor OFF. When ONis selected, the test results
will be displayed at the upper right corner of the screen.
Press Stop On Outp to select ONor OFF.
ON: when failed waveforms are detected, the oscilloscope will stop the test and
enter the STOPstate. At this point, the results of the test remain the same on
the screen (if display is turned on) and only one pulse is output from the [Trig
Out/Calibration] (if enabled) connector at the rear panel.
OFF: the oscilloscope will continue with the test even though failed waveforms
are detected. The test results on the screen will update continuously and a pulse
will be output from the [Trig Out/Calibration] connector at the rear panel
each time a failed waveform is detected.
Press Output to select “Fail” or “Fail+ ”.
Fail: when failed waveforms are detected, there are display and output but the
beeper does not sound.
Fail+
: when failed waveforms are detected, there are display and output and
the beeper sounds (not related to the on/off state of the sound).
Press AuxOutput to quickly turn ONor OFFthe output of test results from the
[Trig Out/Calibration] connector at the rear panel. You can also press Utility
AuxOutput and select PassFailto set this output.