Data Sheet

RIGOL 5 To Trigger the Oscilloscope
MSO4000/DS4000 Users Guide
5-22
Pattern Trigger
This trigger mode will identify a trigger condition by looking for a specified pattern.
This pattern is a logical ANDcombination of the channels. Each channel can have a
value of high (H), low (L) or dont care (X). A rising or falling edge can be specified
for one channel included in the pattern. When an edge is specified, the oscilloscope
will trigger at the edge specified if the pattern set for the other channels are true (H
or L). If no edge is specified, the oscilloscope will trigger on the last edge that makes
the pattern true. If all the channels in the pattern are set to “Dont Care, the
oscilloscope will not trigger.
Trigger Type:
Press Type to select Pattern. At this point, the trigger setting information as shown
in the figure below is displayed at the upper right corner of the screen.
Channel Selection:
Press Source to select CH1 to CH4, EXT or D0 to D15 as the Trigger Source for H, L,
X or edge condition. The current trigger source is displayed at the upper right corner
of the screen.
Pattern Setting:
Press Code to set the pattern code of the current channel. At this point, the pattern
setting area (as shown in the figure below) is displayed at the bottom of the screen.
: set the pattern of the channel selected to “H”, namely the voltage level is
higher than the threshold level of the channel.
: set the pattern of the channel selected to “L, namely the voltage level is
lower than the threshold level of the channel.
: set the pattern of the channel selected to “Dont Care, namely this channel is
not used as a part of the pattern. When all the channels in the pattern are set to
Dont Care, the oscilloscope will not trigger.
or : set the pattern to the rising or falling edge of the channel selected.
Note: Only one rising or falling edge can be specified in the pattern. If one edge item
is currently defined and then another edge item is defined in another channel in the