11 Waveform Record RIGOL
11-9
5. Previous
After the waveform analysis finishes, pressing Previous can locate the error
frame previous to the current error frame. Pressing RUN/STOP can also
perform the operation.
6. Next
After the waveform analysis finishes, pressing Next can locate the error frame
following the current error frame. Pressing SINGLE can also perform the
operation.
Note: You can use the navigation knob to view each frame of waveform in the
analysis.
7. Current Frame
Press Current Frame to adjust the data frame currently displayed and the
adjustable range is from 1 to the total number of frames in this waveform
analysis. Using the navigation knob can also perform the operation.
8. Setup
Press Setup to open the more detailed setting menu.
Screen Start: press this key to set the start point of waveform analysis. The
range is from 5 to 685. The start point must be less than the “Screen End”
currently set.
Screen End: press this key to set the end point of waveform analysis. The
range is from 15 to 695. The end point must be greater than the “Screen
Start” currently set.
Start Frame: press this key to set the start frame of waveform analysis. The
default is the first frame.
End Frame: press this key to set the end frame of waveform analysis. The
default is the last frame.
Threshold: press this key to set the threshold of waveform analysis and the
range is from 1% to 99%. The threshold is used to judge whether the frame
is an error frame. A frame is recognize as an error one if the (relative)
difference value between the frame and the template is equal to or larger
than the threshold currently set.
9. TemplateDisp
Press TemplateDisp to enable or disable template display. For the template