Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may increase measurement error. The probe compensation
procedures are as follows:
1. Perform steps 1, 2, 3 and 4 of
Function Inspectionin the previous section.
2. Check the waveforms displayed and compare them with the following.
Figure 1-9 Probe Compensation
3. Use a nonmetallic driver to adjust the variable capacitor on the probe until the
waveform displayed is as the Perfectly compensatedin the figure above.
Over compensated Perfectly compensated Under compensated