Instructions

RIGOL Chapter 5 To Trigger the Oscilloscope
5-54 MSO7000/DS7000 User's Guide
Trigger Output Connector
The trigger output connector ([TRIG OUT]) on the rear panel of MSO7000/DS7000
series can output trigger signals (hardware trigger) determined by the current
setting.
Figure 5-23 Trigger Output Connector
Press Utility System Aux OUT to select "TrigOut”. A signal which reflects the
current capture rate of the oscilloscope can be output from the [TRIG OUT]
connector each time a trigger is generated by the oscilloscope. If this signal is
connected to a waveform display device to measure the frequency, the measurement
result is the same as the current capture rate.
Note: If you press Utility System Aux OUT to select "PassFail", then in the
pass/fail test, when a pass/fail event is detected, the oscilloscope outputs a pulse
from the rear-panel [TRIG OUT] connector.
Trigger Output Connector