2 RP1001C User’s Guide
Control/Indicator Description
Current flow symbol: The arrow shows
the positive direction of current flow
.
Zero adjustment: After connecting to
power and fixing the current probe to a
fixed position, rotate the knob to adjust
the probe output tothe proper DC offset. It
may also be used to offset a DC signal
component. Zeroing is not needed for AC
measurements unless your instrument
cannot isolate a DC component (if
present).
Range switch: Slide the switch from OFF
to either the 10 mV/A or 100mV/A range.
When 10 mV/A or 100mV/A is selected,
the probe is turned on, and the green
battery indicator (ON) lights..
Battery indicator: The green
battery/power indicator lights when the
probe is turned on. For more information,
please refer to To Install the Battery.
Overload indicator:
The red overload
indicator lights and flicks continuously if
the measured signal is greater than the
maximum range of the selected range
capacity.