User Manual

RIGOL
RP1001C Users Guide 5
Maintenance
Battery
1 The RP1001C current probe uses a single square 9 V battery.
This instrument is a high power product. Please use the
specified alkaline battery.
2 As the battery in the RP1001C current probe is drained,
significant measurement errors may occur. The green battery
indicator will continue to light until a low battery voltage of
6.5 V is reached.
To Install the Battery
1 Remove the test clamp from the circuit.
2 Remove the three screws on the rear cover and take off the
rear cover. You can see the location of the battery and remove
the battery.
3 While observing polarity, install the new
alkaline battery at the
corresponding location.
4 Install the rear cover.
Battery