RIGOL User’s Guide RPL1116 RPL1116 Active Logic Probe © 2013 RIGOL Technologies, Inc. All Rights Reserved.
General Safety Summary Connect and disconnect the probe properly. Observe all terminals ratings. Do not touch exposed connections and components after power on. Do not operate with suspected failures. Do not operate without covers. Do not operate in an explosive atmosphere. Do not operate in wet conditions. Keep product surface clean and dry. Pay attention to handling safety.
Product Overview Being a high-performance active logic probe, RPL1116 connects the digital signals under test to the MSO1000Z series digital oscilloscope to realize the logic analyzer function. The 16 digital channels (D0-D15) of RPL1116 are divided into two channel groups (D0-D7 and D8-D15) each of which includes signal interfaces and ground interfaces. All the channels are marked with different numbers on the label of the probe head to identify different channels.
The Using Method of the Logic Probe 1. Connect RPL1116 to the oscilloscope: connect the active logic head to the main cable input, and then connect the main cable output to the digital signal input terminal at the front panel of the oscilloscope as shown in the figure below.
2. Connect the signals under test to RPL1116: users can connect any number (≤16) of the signals under test to RPL1116 according to the test need. Note that the amplitude of the input signal should not exceed the maximum working voltage range of the probe. RPL1116 provides two connection methods to realize convenient and flexible detection. Method one: users can connect the signals under test through the probe leads separately.
shown in Figure 2. Figure 1 Figure 2 3. Set the probe: press LA at the front panel of the oscilloscope to enter the probe setting menu. Users can view and set the following parameters under this menu: threshold level (the threshold levels of D0-D7 and D8-D15 can be adjusted independently), waveform size (applicable to all the channels; wherein, item L is only available when the number of active channels is no more than 8), channel label, probe calibration and so on.
time or the temperature change is more than 5 degrees, you are recommended to calibrate the probe zero using the auto-calibration function in the LA menu and please disconnect all the connections to the RPL1116 input terminal during the calibration. 4. Function Check: after finishing the above operations, the signal under test will be displayed on the corresponding digital channel on the oscilloscope screen.
Probe Specifications Input channels 16 Threshold range ±15 V Threshold accuracy ±100m V+3% of threshold setting Max input voltage ±40 V Max input dynamic range ±10 V + threshold setting Min voltage swing 500 mV Min detectable pulse width 10 ns Input impedance 100 kΩ±2% Input capacitance About 8 pF Cable length About 90 cm Lead length About 25 cm Operation environment 0℃~50℃, 0~80%RH Storage environment -20℃~60℃, 0~90%RH -7-
Accessories Item Description Quantity 1 Main Cable 1 2 Active Logic Head 1 3 Lead 32 (2×16) 4 Grabber 32 5 Chinese and English User’s Guide 1 6 RPL1116 Packing Box 1 -8-
Accessories Sketch Map Main Cable Active Logic Head Lead Grabber -9-
Contact Us If you have any problem or requirement when using our products or this manual, please contact RIGOL Technologies, Inc. E-mail:service@rigol.com Website:www.rigol.com UGE26X02-1110 Aug.