RIGOL User’s Guide RPL2316 LOGIC PROBE © 2013 RIGOL Technologies, Inc. All Rights Reserved.
General Safety Summary Connect and disconnect the probe properly. Connect the adaptor before using. Observe all terminals ratings. Do not touch exposed connections and components after power on. Do not operate with suspected failures. Do not operate without covers. Do not operate in an explosive atmosphere. Do not operate in wet conditions. Keep product surface clean and dry. Pay attention to handling safety.
Product Overview Being a high performance logic probe, RPL2316 connects the digital buses and signals under test to the MSO4000 series digital oscilloscope. RPL2316 separates 16 digital channels into two branch headers, called channel group. Each header includes 8 signal channels (D0-D7 and D8-D15) plus 12 ground channels. All the channels are marked with different colors and numbers (or letters) on the probe adaptor label to identify different channels.
The Using Method of the Logic Probe 1. Connect RPL2316 to the oscilloscope: push the buttons on both sides of the logic probe single head and connect it (with “RIGOL” facing upward and the label facing downward) to the digital channel input terminal [LOGIC D0-D15] at the front panel of the MSO4000 series digital oscilloscope in the arrow direction as shown in the figure below.
2. Connect the signals under test to RPL2316: users can connect any number (≤16) of the signals under test to RPL2316 in according to the test need. Make sure that the amplitude of the input signal should not exceed the maximum working voltage range while connecting. RPL2316 provides three connection methods with the signals under test for different applications. Method one: you can connect RPL2316 to the two rows of pins on the device under test directly.
Figure 1 Figure 2 Method three: on the basis of method two, you can connect a grabber to each lead and connect it to the device under test as shown in figure 3.
3. Set the probe: press LA at the front panel of the oscilloscope to enter the probe setting menu.
Probe Specifications Input Channels 16 Threshold Range ± 20V Threshold Accuracy ± 100mV+3% of threshold setting Max Voltage Swing ± 40V Min Voltage Swing 500mV Min Detectable Pulse Width 5ns Input Impedance 101kΩ Probe Load About 8pF Cable Length About 100cm Lead Length About 15cm Operation Environment 0~50℃, 0~80%RH Storage Environment -20℃~60℃, 0~90%RH -7-
Accessories Item Description Quantity 1 Main Cable 1 2 Logic Probe Adaptor 2 3 Lead 20 (2×10) 4 Grabber 20 5 Chinese and English User’s Guide 1 6 RPL2316 Soft Carrying Bag 1 -8-
Accessories Sketch Map Logic Probe Adaptor Main Cable Lead Grabber -9-
Contact Us If you have any problem or requirement when using our products or this manual, please contact RIGOL Technologies, Inc. E-mail: service@rigol.com Website: www.rigol.com UGE20X03-1110 Aug.