RIGOL Chapter 2 Functions of the Front Panel of GPSA
2-20 RSA3000E User Guide
enabled, press this key to enter continuous sweep mode. In this case, if the
trigger conditions are met, the instrument will sweep continuously.
In continuous sweep mode, the system will send a trigger initialization signal
automatically and enter the trigger condition judgment directly after each sweep
is completed.
Enter continuous sweep mode and
wait for trigger condition
Sweep again and wait for trigger
condition
Select Cont
Single sweep status
Y N
Figure 2-2 Process of Continuous Sweep
Single
Sets the sweep mode to be single. Selects "Single", and then the Single key on the
front panel is illuminated.
In single sweep mode, press Single under Sweep menu or press Single directly on
the front panel to perform one measurement.
Remarks:
If the instrument is in continuous sweep mode and no measurement function is
enabled, press this key to enter single sweep mode. In this case, if the trigger
conditions are met, the instrument will execute the specified number of sweeps.
(Note: In Average, Max Hold, and Min Hold modes, when performing the single
measurement, the instrument will not stop sweeping until the sweep count has
reached N. Wherein, N is determined by the average count of the current
measurement.)
If the instrument is in continuous sweep mode and the measurement function is
enabled, press this key to set the measurement mode to "Single". In this case, if
the trigger conditions are met, the instrument will execute the specified number
of measurements.
If the system is already in single sweep mode, press this key and then the
system will execute the specified number of sweeps (or measurements) once
the trigger conditions are met.
In single sweep mode, trigger initialization (press Sweep Single or send the