Instructions

Chapter 3 Command System for EMI Mode RIGOL
RSA3000E Programming Guide 3-75
[:SENSe]:FSCan:FINal:DETector<n>:LDELta
Syntax
[:SENSe]:FSCan:FINal:DETector<n>:LDELta <limit_Num>
[:SENSe]:FSCan:FINal:DETector<n>:LDELta?
Description
Sets the limit line used by the specified detector in the final measurement.
Queries the limit line used by the specified detector in the final measurement.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3
1
<limit_Num>
Discrete
1|2|3|4|5|6
1
Return Format
The query returns the limit line used by the specified detector.
Example
The following command sets the limit line used by Detector 1 in the final measurement to Limit Line 2.
:SENSe:FSCan:FINal:DETector1:LDELta 2
The following query returns 2.
:SENSe:FSCan:FINal:DETector1:LDELta?
[:SENSe]:FSCan:MARKer:COUPle:METer
Syntax
[:SENSe]:FSCan:MARKer:COUPle:METer OFF|ON|0|1
[:SENSe]:FSCan:MARKer:COUPle:METer?
Description
Enables or disables the coupling function between the meter and the marker.
Queries the status of the coupling function between the meter and the marker.
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
OFF|0
Return Format
The query returns 0 or 1.
Example
The following command enables the coupling function between the meter and the marker.
:SENSe:FSCan:MARKer:COUPle:METer ON or :SENSe:FSCan:MARKer:COUPle:METer 1
The following query returns 1.
:SENSe:FSCan:MARKer:COUPle:METer?