Instructions

RIGOL Chapter 4 Functions of the Front Panel of EMI
4-24 RSA3000E User Guide
5. Mark Signal
Performs the signal marking operation. Marks the checkbox next to the marked
signal.
6. Mark All
Marks all the signals in the signal table.
7. Clear Mark
Unmarks all the signals in the signal table.
8. Delete
Deletes the signal of the selected type.
Selected: deletes the selected signal from the signal table.
All: marks and deletes all the signals in the signal table.
Marked: deletes all the marked signals from the signal table.
Unmarked: deletes all the unmarked signals from the signal table.
9. Signal Zoom
Zooms the coordinate display of the currently selected signal to 10% of the
current span by taking the midspan frequency as the reference.
10. Zoom Out
Returns the coordinate display of the currently selected signal to its previous
display set before the Signal Zoom.
11. Sort Signals
Sorts the signal in the signal table according to "frequency", "detector
amplitude", or "limit delta detector amplitude".
Wherein, detector amplitude includes "Det 1 Amp", "Det 2 Amp", and "Det 3
Amp"; limit delta detector amplitude includes "Det 1 Amp", "Det 2 Amp",
and "Det 3 Amp".
12. Sort Order
Selects the signal to be sorted in ascending or descending order.
13. Auto Sort
Enables or disables the auto sort function. When enabled, the measurement will
sort the signal according to the previously set sort order and sort type.
Otherwise, the signal will be added to the signal table in the descending order of
the trace amplitude.
14. Search Criteria
Sets the search criteria. When the peak that meets the conditions is searched,
the signal will be added to the signal table.
Peak Criteria: performs the search and finds the peak that meets the peak
criteria.