Instructions

Chapter 2 Functions of the Front Panel of GPSA RIGOL
RSA3000E User Guide 2-11
Atten Auto
Selects "Manual" or "Auto" to be the current attenuation mode.
Remarks:
To select "Manual" to be the attenuation mode, press Attenuation to set the
attenuation value, and the attenuation mode will be automatically switched to
"Manual".
When you select "Auto" to be the attenuation mode, the attenuator will
automatically adjust the attenuation value to meet the current amplitude
setting.
RF Preamp
Sets the RF front-end preamplifier to be on or off. When the signal under test is a
low-level signal, turning on the preamplifier can reduce the displayed average noise
level, so that you can distinguish low-level signals from the noise. By default, the
preamp gain is 20 dB.
Y Axis Unit
Sets the unit of the Y-axis to dBm, dBmV, dBuV, Volts, or Watts. Wherein, dBm, dBmV,
and dBuV are for Log scale; Volts and Watts are for Linear scale. The default unit is
dBm.
Remarks:
The conversion relations between units are as follows:
×=
0.001W
1
R
2
Volts
10lgdBm
(2-4)
=
V1
Volts
20lgVdB
µ
µ
(2-5)
=
1mV
Volts
20lgdBmV
(2-6)
R
2
Volts
Watts =
(2-7)
Wherein,
R
denotes the reference resistance.