Instructions

Chapter 2 Functions of the Front Panel of GPSA RIGOL
RSA3000E User Guide 2-35
3. Reference Position
After the normalize function is enabled, you can adjust the vertical position of
Reference Level on the screen by adjusting the reference position.
The function of this menu is similar to that of Reference Level. When it is set
to 0%, Reference Level is displayed at the bottom of the graticule and at
the top line of the graticule when it is set to 100%.
You can use the numeric keys, the knob, and the arrow keys on the front
panel to modify this parameter; also you can modify it on the touchscreen.
For details, refer to descriptions in "Parameter Setting".
Table 2-29 Tracking Generator Reference Position
Parameter
Remarks
Default
100%
Range
0% to 100%
Unit
%
Knob Step
1%
Left/Right Arrow Key Step
Up/Down Arrow Key Step
10%
4. Reference Trace
Sets whether to display the reference trace. If you set "Reference Trace" to be
"On", the reference trace saved (Trace 6) will be shown.
5. Save Ref Trace
Saves the data of Trace1 to Trace6 as the reference value for normalization. This
operation should be done before you enable the normalize function.