RSA3000E Series Real-time Spectrum Analyzer Ultra-Real technology Frequency: up to 3 GHz Displayed average noise level (DANL): <-161 dBm (typical) Phase noise: <-102 dBc/Hz (typical) Level measurement uncertainty: <1.0 dB 3 GHz tracking generator Min.
RSA3000E Series Real-time Spectrum Analyzer Built-in Linux operating system reliable and stable interface Built-in quad-core processor high processing speed 10.1-inch capacitive multi-touch screen supporting touch gestures Support keyboard and mouse operation TG: 100kHz to 1.5/3.0GHz -40 dBm to 0 dBm RF: 9 kHz to 1.5/3.
Specifications Specifications are valid under the following conditions: the instrument is within the calibration period, is stored for at least two hours at 0℃ to 50℃ temperature, and is warmed up for 40 minutes. Unless otherwise noted, the specifications in this manual include the measurement uncertainty. Typical: characteristic performance, which 80 percent of the measurement results will meet at room temperature (approximately 25℃).
GPSA Mode Frequency Frequency Readout Accuracy Marker Frequency Resolution span/(number of sweep points - 1) Marker Frequency Uncertainty ±(marker frequency readout × reference frequency accuracy + 1% × span + 10% × resolution bandwidth + marker frequency resolution) Frequency Counter Resolution 1 Hz Uncertainty ±(marker frequency readout × reference frequency accuracy + counter resolution) Frequency Span Range 0 Hz, 10 Hz to maximum frequency Resolution 2 Hz Uncertainty ±span/(number of sweep
Amplitude Measurement Range Range fc ≥ 10 MHz DANL to +30 dBm Maximum Safe Input Level[1] DC Voltage CW RF Power 50 V +30 dBm, attenuation ≥ 40 dB, preamp off. -10 dBm, attenuation = 20 dB, preamp on. Maximum Damage Level CW RF Power +33 dBm (2 W) Displayed Average Noise Level (DANL) attenuation = 0 dB, sample detector, trace averages ≥ 50, tracking generator off, normalized to 1 Hz, 20℃ to 30℃, input impedance = 50 Ω.
Level Display Logarithmic Scale 1 dB to 200 dB Linear Scale 0 to reference level Number of Display Points 801 Number of Traces 6 Trace Detector normal, pos-peak, neg-peak, sample, RMS average, voltage average, and quasi-peak (Option RSA3000E-EMC) Trace Function clear write, max hold, min hold, average, view, blank Scale Unit dBm, dBmV, dBμV, nV, μV, mV, V, nW, μW, mW, W Frequency Response attenuation = 10 dB, relative to 50 MHz, 20℃ to 30℃ Preamp off 100 kHz to 3.
Input Attenuation Switching Uncertainty Setting Range Switching Uncertainty 0 dB to 50 dB, in 1 dB step fc = 50 MHz, relative to 10 dB, preamp off, 20℃ to 30℃ <0.3 dB Absolute Amplitude Accuracy fc = 50 MHz, peak detector, preamp off, attenuation = 10 dB, input signal level = -10 dBm, 20℃ to 30℃ Uncertainty <0.3 dB Reference Level Range Logarithmic Scale -170 dBm to +30 dBm, in 0.01 dB step Linear Scale 707 pV to 7.07 V, 0.11% (0.
RF Input VSWR attenuation ≥10 dB, preamp off VSWR 300 kHz to 3.0 GHz <1.6 (nominal) Distortion Second Harmonic Intercept (SHI) Third-order Intercept (TOI) fc ≥ 50 MHz, input signal level = -20 dBm, attenuation = 0 dB, preamp off. +45 dBm fc ≥ 50 MHz, two -20 dBm tones at input mixer spaced by 200 kHz, attenuation = 0 dB, preamp off.
Sweep Sweep Sweep Time Sweep Time Uncertainty span ≥ 10 Hz 1 ms to 4,000 s zero span 1 μs to 6,000 s span ≥ 10 Hz, RBW ≥ 1 kHz 5% (nominal) zero span (sweep time > 1 ms) 5% (nominal) Sweep Mode continue, single Trigger Trigger Trigger Source Trigger Delay free run, external 1, external 2, video span ≥ 10 Hz 0 to 500 ms zero span 0 to 500 ms Tracking Generator Tracking Generator Output Frequency Range RSA3015E-TG 100 kHz to 1.5 GHz RSA3030E-TG 100 kHz to 3.
RTSA Mode Real-time Analysis Bandwidth 10 MHz Min. Signal Duration for 100% POI at the Full-Scale Accuracy maximum span, default Kaiser Window Trace Detector pos-peak, neg-peak, sample, average Number of Traces 6 Window Type Hanning, Blackman-Harris, Rectangular, Flattop, Kaiser, and Gaussian 9.3 μs provides 6 RBWs for each window, except the Rectangular; for Kaiser window Resolution Bandwidth Span Min. bandwidth Max. bandwidth 10 MHz 25.1 kHz 804 kHz 1 MHz 2.51 kHz 80.
VSA Mode (Option RSA3000E-ASK/FSK) Capture Oversampling Capture Oversampling Capture Length Capture Oversampling = 4 Capture Oversampling = 8 Capture Oversampling = 16 Sample Rate Maximum Sample Rate Symbol Rate 4, 8, 16 Maximum 4096 Maximum 2048 Maximum 1024 12.
General Specifications Display Type capacitive multi-touch screen Resolution 1024 × 600 pixels Size 10.1'' Color 24-bit color Printer Supported Protocol network printer Mass Memory Mass Memory Internal Storage 512 MB (nominal) External Storage USB storage device (not supplied) Power Input Voltage Range, AC 100 V to 240 V (nominal) AC Frequency 45 Hz to 440 Hz Power Consumption 55 W (typical), max.
Input/Output Front Panel Connector RF Input TG Output Impedance 50 Ω (nominal) Connector N-type female Impedance 50 Ω (nominal) Connector N-type female Frequency 10 MHz Output Level +3 dBm to +10 dBm, +7 dBm (typical) Impedance 50 Ω (nominal) Connector BNC female Frequency 10 MHz ± 5 ppm Input Level 0 dBm to +10 dBm Impedance 50 Ω (nominal) Connector BNC female Impedance ≥1 kΩ (nominal) Connector BNC female Level 5 V TTL level Internal/External Reference Internal Reference E
Order Information Model Standard Accessories Option Optional Accessories Description Order No. Real-time Spectrum Analyzer, 9 kHz to 1.5 GHz RSA3015E Real-time Spectrum Analyzer, 9 kHz to 3 GHz RSA3030E Real-time Spectrum Analyzer, 9 kHz to 1.
HEADQUARTER EUROPE NORTH AMERICA JAPAN RIGOL TECHNOLOGIES, INC. No.8 Keling Road, New District,Suzhou, JiangSu,P.R.China Tel:+86-400620002 Email:info@rigol.com RIGOL TECHNOLOGIES EU GmbH Lindbergh str. 4 82178 Puchheim Germany Tel: 0049-89/89418950 Email: info-europe@rigol.com RIGOL TECHNOLOGIES, USA INC. 8140 SW Nimbus Ave. Beaverton, OR 97008 Tel: 877-4-RIGOL-1 Fax: 877-4-RIGOL-1 Email: info@rigol.com RIGOL TECHNOLOGIES JAPAN, LLC MJ Bldg.