Instructions

Chapter 2 Command Systems for GPSA and RTSA Modes RIGOL
RSA3000E Programming Guide 2-37
:CALCulate:MARKer<n>:Y
Syntax
:CALCulate:MARKer<n>:Y <amp>
:CALCulate:MARKer<n>:Y?
Description
Sets the Y-axis value of the specified fixed marker.
Queries the Y-axis value of the specified marker, and its default unit is dBm.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8
——
<amp>
Consecutive Real Number
-170 dBm to 30 dBm
——
Remarks
This setting command is only available when there is only one marker and it is a fixed marker.
If the marker mode of the specified marker is Position or Fixed, the query command queries the Y value of
the marker
.
If the marker mode of the specified marker is Delta, the query command queries the Y-axis difference
between the reference marker and the Delta marker.
The PvT only supports setting the marker on Trace 1.
Return Format
The query returns the Y-axis value in scientific notation.
Example
The following command sets the Y-axis value of Marker 1 is -59.6 dBm.
:CALCulate:MARKer1:Y -59.6
The following query returns -5.960000000e+01.
:CALCulate:MARKer1:Y?
:CALCulate:MARKer<n>:Z:POSition
Syntax
:CALCulate:MARKer<n>:Z:POSition <integer>
:CALCulate:MARKer<n>:Z:POSition?
Description
Sets the trace number of the trace where the marker stays in the Spectrogram view.
Queries the trace number of the trace where the marker stays in the Spectrogram view.
Parameter
Name
Type
Range
Default
<n>
Discrete
1
|
2|3
|
4|5|6|7|8
——
<integer>
Integer
1 to 8192
1
Remarks
Only when you select the Spectrogram view of the Spectrogram, Density Spectrogram or PvT Spectrogram
in RTSA mode, can this command be effective.
Return Format
The query returns the trace number in integer.