Instructions

Chapter 2 Functions of the Front Panel of GPSA RIGOL
RSA5000 User's Guide 2-7
Start
Execute a
sweep
Search for
peak and
mark it as
Marker
Active Marker
exists?
N
Set marker
f
requency to
center freq
Y
Execute the
next sweep
Search for and mark the
frequency p
oint (variation
< 3
dB
) n
ear the Marker
Figure 2-1 Signal Tracking Process
Remarks:
If an active marker currently exists, when Signal Track is enabled, the
instrument will search for and mark the point (whose amplitude variation is less
than 3 dB) near the marker. Then, set the frequency at this point to be the
center frequency to keep the signal at the center of the screen.
If no marker is currently active, when Signal Track is enabled, the instrument will
activate Marker1, execute a peak search automatically, and set the frequency
value at the current peak to be the center frequency to keep the signal always
displayed at the center of the screen.
In continuous sweep, the system tracks the signal continuously; in single sweep,
the instrument only performs a single signal track.
The signal tracking function is only applicable to the swept analysis. The signal
tracking function is invalid in the following conditions:
In zero span mode;
When the tracking generator is enabled;
When the trace is not updated;
When the continuous peak search is enabled;
In advanced measurement mode.