Instructions

Chapter 2 Functions of the Front Panel of GPSA RIGOL
RSA5000 User's Guide 2-13
Max Mixer Lvl
Sets the maximum input level of the mixer according to the amplitude of the signal.
Remarks:
For the high-level input signal, select a smaller maximum mixer level to increase
the input attenuation and reduce the distortion of the signal; for the low-level
input signal, select a larger maximum mixer level to reduce the input attenuation
and noise.
You can use the numeric keys, the knob, and the arrow keys on the front panel
to modify this parameter; also you can modify it on the touchscreen. For details,
refer to descriptions in "Parameter Setting".
Table 2-10 Maximum Mixing Level
Parameter
Remarks
Default
-10 dBm
Range
-50 dBm to -10 dBm
Unit
dBm, -dBm, mV, uV
Knob Step
1 dBm
Left/Right Arrow Key Step
Up/Down Arrow Key Step
5 dBm
Ref Offset
Adds an offset value to the reference level to compensate for gains or losses
generated between the DUT and the analyzer input.
Remarks:
The offset value does not affect the trace position, but will modify the reference
level readout and the marker amplitude readout.
You can use the numeric keys, the knob, and the arrow keys on the front panel
to modify this parameter; also you can modify it on the touchscreen. For details,
refer to descriptions in "Parameter Setting".
Table 2-11 Reference Level Offset
Parameter
Remarks
Default
0 dB
Range
-300 dB to 300 dB
Unit
dB
Knob Step
1 dB
Left/Right Arrow Key Step
Up/Down Arrow Key Step
5 dB