Instructions

Table Of Contents
RIGOL Chapter 2 Command System
2-100 RSA5000 Programming Guide
:READ:ACPower?
Syntax
:READ:ACPower?
Description
Executes one adjacent channel power measurement and returns the measurement results.
Remarks
The power unit of the returned value is consistent with the current Y-axis unit.
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
This command is only available for GPSA mode.
Return Format
The query returns 5 values (main channel power, upper channel power, the power difference between the
upper channel and the main channel (in dBc), the lower channel power, the power difference between the
lower channel and the main channel (in dBc)) in scientific notation, and the values are separated by
commas. For example,
-5.150423000e+01,-5.173441000e+01,-2.301865000e-01,-5.142665000e+01,7.757568000e-02
:READ:ACPower:LOWer?
Syntax
:READ:ACPower:LOWer?
Description
Executes one adjacent channel power measurement and returns the lower channel power.
Remarks
The power unit of the return value is consistent with the current Y-axis unit.
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
This command is only available for GPSA mode.
Return Format
The query returns the lower channel power in scientific notation. For example, -5.142665000e+01