:READ:CNRatio:CARRier?
Syntax
:READ:CNRatio:CARRier?
Description
Executes one C/N ratio measurement and returns the carrier power.
Remarks
The power unit of the return value is consistent with the current Y-axis unit.
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
This command is only available for GPSA mode.
Return Format
The query returns the carrier power in scientific notation. For example, -1.484203000e+01
:READ:CNRatio:CNRatio?
Syntax
:READ:CNRatio:CNRatio?
Description
Executes one C/N ratio measurement and returns the C/N ratio.
Remarks
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
This command is only available for GPSA mode.
Return Format
The query returns the C/N ratio in scientific notation. For example, 8.956909000e-02
:READ:CNRatio:NOISe?
Syntax
:READ:CNRatio:NOISe?
Description
Executes one C/N ratio measurement and returns the noise power.
Remarks
The power unit of the return value is consistent with the current Y-axis unit.
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
This command is only available for GPSA mode.
Return Format