Chapter 2 Command System RIGOL
RSA5000 Programming Guide 2-107
:READ:OBWidth:OBWidth:FERRor?
Syntax
:READ:OBWidth:OBWidth:FERRor?
Description
Executes one occupied bandwidth measurement and returns the transmit frequency error.
Remarks
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
This command is only available for GPSA mode.
Return Format
The query returns the transmit frequency error in scientific notation. The unit is Hz. For example,
2.000000000+04
:READ:SANalyzer<n>?
Syntax
:READ:SANalyzer<n>?
Description
Queries the measurement results or trace data in the buffer.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6
——
Remarks
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
This command is only available for GPSA mode.
Return Format
The query returns the measurement data of Trace n in scientific notation. The measurement data are
returned with a pair of data (x,y), separated by commas. The unit is Hz.
Example
The following query returns
0,0,0,0,-1.000000000e+02,0,801,0,0,0,-3.19e+01,0.00e+00,0.00e+00,0.00e+00,0.00e+00,0.00e+00,0.
00e+00,0.00e+00
:FETCh:SANalyzer1?