Instructions

Table Of Contents
RIGOL Chapter 2 Command System
2-108 RSA5000 Programming Guide
:READ:TOIntercept?
Syntax
:READ:TOIntercept?
Description
Executes one TOI measurement and returns the measurement results.
Remarks
The amplitude unit of the return value is consistent with the current Y-axis unit.
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
This command is only available for GPSA mode.
Return Format
The query returns the measurement results of TOI in the following format:
Base Lower frequency (Hz), amplitude, Base Upper frequency (Hz), amplitude, 3rd Order Lower frequency
(Hz), amplitude, intercept, 3rd Order Upper frequency (Hz), amplitude, intercept.
For example,
1.500450000e+09,-8.131735000e+01,1.500450000e+09,-8.131735000e+01,1.500450000e+09,-8.13173
5000e+01,-8.131735000e+01,1.500450000+09,-8.131735000e+01,-8.131735000e+01
:READ:TOIntercept:IP3?
Syntax
:READ:TOIntercept:IP3?
Description
Executes one TOI measurement and returns the value of the intercept of the Third Order Lower or the
Third Order Upper, whichever is smaller.
Remarks
This command will select data from the latest measurement results, and then transmit the data to the
output buffer.
This command is only available for GPSA mode.
Return Format
The query returns the minor one in scientific notation. For example, -8.131735000e+01