Instructions

Table Of Contents
RIGOL Chapter 2 Command System
2-114 RSA5000 Programming Guide
[:SENSe]:ACPower:AVERage:TCONtrol
Syntax
[:SENSe]:ACPower:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:ACPower:AVERage:TCONtrol?
Description
Selects the average mode of the adjacent channel power measurement.
Queries the average mode of the adjacent channel power measurement.
Parameter
Name
Type
Range
Default
——
Keyword
EXPonential|REPeat
EXPonential
Remarks
EXPonential: indicates the exponential average.
REPeat: indicates the repeat average.
When "EXPonential" is selected, the result is the exponential average of the measurement results obtained
in the past N times (N is specified in
[:SENSe]:ACPower:AVERage:COUNt).
When "REPeat" is selected, the result is the arithmetic average of the measurement results obtained in the
past N times (N is specified in
[:SENSe]:ACPower:AVERage:COUNt).
This command is only valid when the adjacent channel power measurement in GPSA mode is enabled.
Return Format
The query returns EXP or REP
.
Example
The following command selects repeat average to the average mode.
:SENSe:ACPower:AVERage:TCONtrol REPeat
The following query returns REP.
:SENSe:ACPower:AVERage:TCONtrol?
[:SENSe]:ACPower:BANDwidth:ACHannel
Syntax
[:SENSe]:ACPower:BANDwidth:ACHannel <freq>
[:SENSe]:ACPower:BANDwidth:ACHannel?
Description
Sets the bandwidth of the adjacent channel.
Queries the bandwidth of the adjacent channel.
Parameter
Name
Type
Range
Default
<freq>
Consecutive Real Number
3 Hz to 2.166666 GHz
2 MHz
Remarks
This command is only valid when the adjacent channel power measurement in GPSA mode is enabled.
Return Format
The query returns the bandwidth of the adjacent channel in scientific notation. Its unit is Hz.