Instructions

Table Of Contents
RIGOL Chapter 2 Command System
2-138 RSA5000 Programming Guide
whichever is the maximum.
When Max Hold is disabled, the current measurement result is displayed.
Max Hold and average measurement are mutually exclusive. When Max Hold is enabled, the average
measurement will be automatically disabled.
Return Format
The query returns 0 or 1.
Example
The following command enables the Max Hold.
:SENSe:EBWidth:MAXHold:STATe ON or :SENSe:EBWidth:MAXHold:STATe 1
The following query returns 1.
:SENSe:EBWidth:MAXHold:STATe?
[:SENSe]:EBWidth:XDB
Syntax
[:SENSe]:EBWidth:XDB <real_amp>
[:SENSe]:EBWidth:XDB?
Description
Sets the value of X dB for the EBW measurement.
Queries the value of X dB for the EBW measurement.
Parameter
Name
Type
Range
Default
<real_amp>
Consecutive Real Number
-100 dB to -0.1 dB
-10 dB
Remarks
This command is only valid when the emission bandwidth measurement in GPSA mode is enabled.
Return Format
The query returns the X dB value in scientific notation.
Example
The following command sets the X dB value to -20 dB.
:SENSe:EBWidth:XDB -20
The following query returns -2.000000000e+01.
:SENSe:EBWidth:XDB?