RIGOL Chapter 2 Command System
2-146 RSA5000 Programming Guide
Example
The following command sets the average mode to Repeat.
:SENSe:HDISt:AVERage:TCONtrol REPeat
The following query returns REP.
:SENSe:HDISt:AVERage:TCONtrol?
[:SENSe]:HDISt:NUMBers
Syntax
[:SENSe]:HDISt:NUMBers <integer>
[:SENSe]:HDISt:NUMBers?
Description
Sets the number of harmonics to be measured.
Queries the number of harmonics to be measured.
Parameter
Name
Type
Range
Default
<integer>
Integer
2 to 10
10
Remarks
This command is only valid when the harmonic distortion measurement in GPSA mode is enabled.
Return Format
The query returns the number of harmonics in integer.
Example
The following command sets the number of harmonics to 5.
:SENSe:HDISt:NUMBers 5
The following query returns 5.
:SENSe:HDISt:NUMBers?
[:SENSe]:HDISt:TIME
Syntax
[:SENSe]:HDISt:TIME <time>
[:SENSe]:HDISt:TIME?
Description
Sets the sweep time of the harmonic distortion measurement.
Queries the sweep time of the harmonic distortion measurement.
Parameter
Name
Type
Range
Default
<time>
Consecutive Real Number
20 μs to 6 ks
1 ms
Remarks
This command is only valid when the harmonic distortion measurement in GPSA mode is enabled.
This setting changes the sweep time of the analyzer.