Instructions

Table Of Contents
Chapter 2 Command System RIGOL
RSA5000 Programming Guide 2-153
Return Format
The query returns 0 or 1.
Example
The following command disables the auto setting mode of the input attenuation.
:SENSe:POWer:RF:ATTenuation:AUTO OFF or :SENSe:POWer:RF:ATTenuation:AUTO 0
The following query returns 0.
:SENSe:POWer:RF:ATTenuation:AUTO?
[:SENSe]:POWer[:RF]:GAIN[:STATe]
Syntax
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Description
Enable or disable the preamplifier.
Queries the status of the preamplifier.
Parameter
Name
Type
Range
Default
——
Bool
OFF|ON|0|1
OFF|0
Return Format
The query returns 0 or 1.
Example
The following command enables the preamplifier.
:SENSe:POWer:RF:GAIN:STATe ON or :SENSe:POWer:RF:GAIN:STATe 1
The following query returns 1.
:SENSe:POWer:RF:GAIN:STATe?
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]
Syntax
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <ampl>
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?
Description
Sets the maximum power of the input mixer.
Queries the maximum power of the input mixer.
Parameter
Name
Type
Range
Default
<ampl>
Consecutive Real Number
-50 dBm to -10 dBm
-10 dBm
Return Format
The query returns the maximum power of the input mixer in scientific notation. The unit is dBm.