Instructions

Table Of Contents
RIGOL Chapter 2 Command System
2-154 RSA5000 Programming Guide
Example
The following command sets the maximum power of the input mixer to -20 dBm.
:SENSe:POWer:RF:MIXer:RANGe:UPPer -20
The following query returns -2.000000000e+01.
:SENSe:POWer:RF:MIXer:RANGe:UPPer?
[:SENSe]:SIGCapture:2FSK:AMPDown
Syntax
[:SENSe]:SIGCapture:2FSK:AMPDown <ample>
[:SENSe]:SIGCapture:2FSK:AMPDown?
Description
Sets the lower limit of the amplitude of 2FSK signal.
Queries the lower limit of the amplitude of 2FSK signal.
Parameter
Name
Type
Range
Default
<ample>
Consecutive Real Number
-400 dBm to amplitude upper limit
-100 dBm
Remarks
This command is only available for RTSA mode and valid when the SSC function is enabled.
Return Format
The query returns the amplitude lower limit in scientific notation.
Example
The following command sets the amplitude lower limit to -20 dBm.
:SENSe:SIGCapture:2FSK:AMPDown -20
The following query returns -2.000000000e+01.
:SENSe:SIGCapture:2FSK:AMPDown?
[:SENSe]:SIGCapture:2FSK:AMPUp
Syntax
[:SENSe]:SIGCapture:2FSK:AMPUp <ample>
[:SENSe]:SIGCapture:2FSK:AMPUp?
Description
Sets the upper limit of the amplitude of 2FSK signal.
Queries the upper limit of the amplitude of 2FSK signal.
Parameter
Name
Type
Range
Default
<ample>
Consecutive Real Number
-100 dBm to 320 dBm
-10 dBm
Remarks
This command is only available for RTSA mode and valid when the SSC function is enabled.