Instructions

Table Of Contents
RIGOL Chapter 2 Command System
2-160 RSA5000 Programming Guide
[:SENSe]:SIGCapture:2FSK:RESet
Syntax
[:SENSe]:SIGCapture:2FSK:RESet
Description
Performs the reset operation, deletes the original signal, and restarts to capture the signal.
Remarks
This command is only available for RTSA mode and valid when the SSC function is enabled.
[:SENSe]:SIGCapture:2FSK:SIGNal
Syntax
[:SENSe]:SIGCapture:2FSK:SIGNal 0|1|2
[:SENSe]:SIGCapture:2FSK:SIGNal?
Description
Selects the signal whose Limit value is required to be modified in the pass/fail test.
Queries the signal whose Limit value is required to be modified in the pass/fail test.
Parameter
Name
Type
Range
Default
--
Discrete
0|1|2
0
Remarks
0: selects to modify Signal 1.
1: selects to modify Signal 2.
2: selects to modify Signal 3.
This command is only available for RTSA mode and valid when the SSC function is enabled.
Return Format
The query returns 0, 1, or 2.
Example
The following command selects to modify Signal 1.
:SENSe:SIGCapture:2FSK:SIGNal 0
The following query returns 0.
:SENSe:SIGCapture:2FSK:SIGNal?