Instructions

Table Of Contents
RIGOL Chapter 2 Command System
2-164 RSA5000 Programming Guide
[:SENSe]:TOI:AVERage:TCONtrol
Syntax
[:SENSe]:TOI:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:TOI:AVERage:TCONtrol?
Description
Selects the average mode of the TOI measurement.
Queries the average mode of the TOI measurement.
Parameter
Name
Type
Range
Default
——
Keyword
EXPonential|REPeat
EXPonential
Remarks
EXPonential: indicates the exponential average.
REPeat: indicates the repeat average.
When "EXPonential" is selected, the result is the exponential average of the measurement results obtained
in the past N times (N is specified in
[:SENSe]:TOI:AVERage:COUNt).
When "REPeat" is selected, the result is the arithmetic average of the measurement results obtained in the
past N times (N is specified in
[:SENSe]:TOI:AVERage:COUNt).
This command is only valid when the TOI measurement in GPSA mode is enabled.
Return Format
The query returns EXP or REP
.
Example
The following command sets the average mode to Repeat.
:SENSe:TOI:AVERage:TCONtrol REPeat
The following query returns REP.
:SENSe:TOI:AVERage:TCONtrol?
[:SENSe]:TOI:FREQuency:SPAN
Syntax
[:SENSe]:TOI:FREQuency:SPAN <freq>
[:SENSe]:TOI:FREQuency:SPAN?
Description
Sets the span of the TOI measurement.
Queries the span of the TOI measurement.
Parameter
Name
Type
Range
Default
<freq>
Consecutive Real Number
10 Hz to 6.5 GHz
2 MHz
Remarks
This command is only valid when the TOI measurement in GPSA mode is enabled.
This setting will change the span of the analyzer.
Return Format
The query returns the span in scientific notation. The unit is Hz.