Instructions

Table Of Contents
RIGOL Chapter 2 Command System
2-188 RSA5000 Programming Guide
:SYSTem:PRESet:TYPE
Syntax
:SYSTem:PRESet:TYPe FACTory|USER1|USER2|USER3|USER4|USER5|USER6
:SYSTem:PRESet:TYPe?
Description
Selects the preset type of the system to be the factory mode or any one of the items from User1 to User6.
Queries the preset type of the system.
Parameter
Name
Type
Range
Default
——
Keyword
FACTory|USER1|USER2|USER3|USER4|USER5|USER6
FACTory
Return Format
The query returns FACT, USER1, USER2, USER3, USER4, USER5, or USER6.
Example
The following command sets the preset type of the system to User5.
:SYSTem:PRESet:TYPe USER5
The following query returns USER5.
:SYSTem:PRESet:TYPe?
:SYSTem:PRESet:USER:SAVE
Syntax
:SYSTem:PRESet:USER:SAVE
Description
Saves the current user setting.
Remarks
If the current preset type is "FACTory", save the current user setting to User1. If the current preset type is
"User1 through User6", save the current user setting to the specified user.