Instructions

Table Of Contents
RIGOL Chapter 2 Command System
2-30 RSA5000 Programming Guide
:CALCulate:MARKer<n>[:SET]:CENTer
Syntax
:CALCulate:MARKer<n>[:SET]:CENTer
Description
Sets the frequency of the specified marker to the center frequency of the analyzer.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8
——
Remarks
If the marker mode of the specified marker is Position or Fixed, the center frequency will be set to the
frequency of the marker.
If the specified marker mode is Delta, the center frequency will be set to the frequency of the Delta marker.
This function is invalid in zero span.
This command is invalid for the PvT measurement mode in RTSA.
Example
The following command sets the frequency at Marker 1 (Position) to the center frequency.
:CALCulate:MARKer1:SET:CENTer
:CALCulate:MARKer<n>[:SET]:DELTa:CENTer
Syntax
:CALCulate:MARKer<n>[:SET]:DELTa:CENTer
Description
Sets the frequency difference of the specified Delta marker to the center frequency of the analyzer.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8
——
Remarks
It is only valid when the current marker mode is "Delta".
This function is invalid in zero span.
This command is invalid for the PvT measurement mode in RTSA.
Example
The following command sets the frequency difference of the Delta Marker 1 to the center frequency of the
analyzer.
:CALCulate:MARKer1:SET:DELTa:CENTer