Instructions

Table Of Contents
RIGOL Chapter 2 Command System
2-38 RSA5000 Programming Guide
:CALCulate:MARKer<n>:Z:POSition
Syntax
:CALCulate:MARKer<n>:Z:POSition <integer>
:CALCulate:MARKer<n>:Z:POSition?
Description
Sets the trace number of the trace where the marker stays in the Spectrogram view.
Queries the trace number of the trace where the marker stays in the Spectrogram view.
Parameter
Name
Type
Range
Default
<n>
Discrete
1
|
2|3
|
4|5|6|7|8
——
<integer>
Integer
1 to 8192
1
Remarks
Only when you select the Spectrogram view of the Spectrogram, Density Spectrogram or PvT Spectrogram
in RTSA mode, can this command be effective.
Return Format
The query returns the trace number in integer.
Example
The following command sets the trace number of the trace where the marker1 stays in to 100.
:CALCulate:MARKer1:Z:POSition 100
The following query returns 100.
:CALCulate:MARKer1:Z:POSition?
:CALCulate:MATH
Syntax
:CALCulate:MATH
<trace_destination>,<function_math>,<trace1_operation>,<trace2_operation>,<rel_offset>,<rel_refer
ence>
:CALCulate:MATH? TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6
Description
Sets mathematical operations between traces and, in some cases, user-defined offsets.
Queries the math operation function of the specified trace.
Parameter
Name
Type
Range
Default
<trace_destination> Keyword
TRACE1|TRACE2|TRACE3|TRACE4|TRACE
5|TRACE6
——
<function_math> Keyword
PDIFference|PSUM|LOFFset|LMOFFset|LD
IFference|OFF
OFF
<trace1_operation> Keyword
TRACE1|TRACE2|TRACE3|TRACE4|TRACE
5|TRACE6
TRACE5
<trace2_operation> Keyword
TRACE1|TRACE2|TRACE3|TRACE4|TRACE
5|TRACE6
TRACE6
<rel_offset>
Consecutive Real
Number
-100 dB to 100 dB 0 dB
<rel_reference>
Consecutive Real
Number
-170 dBm to 30 dBm 0 dBm