Chapter 2 Command System RIGOL
RSA5000 Programming Guide 2-43
:CONFigure:DSPEctrogram
Syntax
:CONFigure:DSPEctrogram
Description
Sets the analyzer to be in the Density Spectrogram measurement state.
Remarks
This command is only available for RTSA mode.
Running this command will discontinue the current measurement and restore the specified measurement of
the instrument to the default factory state.
:CONFigure:EBWidth
Syntax
:CONFigure:EBWidth
Description
Sets the analyzer to be in the EBW measurement state.
Remarks
This command is only available for GPSA mode.
Running this command will discontinue the current measurement and restore the specified measurement of
the instrument to the default factory state.
After changing any measurement setup, running the
:READ Commands will initialize the measurement, and
you do not have to set the measurement to the default.
:CONFigure:HDISt
Syntax
:CONFigure:HDISt
Description
Sets the analyzer to be in the harmonic distortion measurement state.
Remarks
This command is only available for GPSA mode.
Running this command will discontinue the current measurement and restore the specified measurement of
the instrument to the default factory state.
After changing any measurement setup, running the
:READ Commands will initialize the measurement, and
you do not have to set the measurement to the default.
:CONFigure:LPSTep
Syntax
:CONFigure:LPSTep
Description
Resets all the parameters in the current measurement mode.