Instructions

Table Of Contents
Chapter 2 Command System RIGOL
RSA5000 Programming Guide 2-65
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel
Syntax
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel <ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
Description
Sets the reference level.
Queries the reference level.
Parameter
Name
Type
Range
Default
<ampl>
Consecutive Real Number
-170 dBm to 30 dBm
0 dBm
Return Format
The query returns the reference level in scientific notation.
Example
The following command sets the reference level to -10 dB.
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel -10
The following query returns -1.000000e+01.
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel?
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet
Syntax
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel_ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet?
Description
Sets the reference level offset.
Queries the reference level offset.
Parameter
Name
Type
Range
Default
<rel_ampl>
Consecutive Real Number
-300 dB to 300 dB
0 dB
Remarks
The offset of the reference level only modifies the reference level and amplitude readout of the marker, but
does not change the position of the curve.
Return Format
The query returns the offset of the reference level in scientific notation.
Example
The following command sets the reference level offset to 10 dB.
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel:OFFSet 10
The following query returns 1.000000e+01.
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel:OFFSet?