Specifications
3 
Vertical 
A/D Converter 
8-bit resolution, all channels sample simultaneously 
Volts/div Range 
2mV/div~10V/div (at the input terminal connecting to BNC) 
Maximum Input 
Maximum input voltage on analog channel   
CAT I    300Vrms, 1000Vpk; instantaneous overvoltage 1000Vpk 
CAT II 100Vrms, 1000Vpk 
RP2200 10:1:CAT II 300Vrms 
RP3200 10:1:CAT II 300Vrms 
RP3300 10:1:CAT II 300Vrms 
Offset Range   
± 40V (250mV/div~10V/div) 
± 2V (2mV/div~245mV/div) 
Analog Bandwidth 
100MHz (DS1102D,DS1102E) 
50MHz (DS1052D, DS1052E) 
Single-shot Bandwidth 
100MHz (DS1102D, DS1102E) 
50MHz (DS1052D, DS1052E) 
Selectable Analog 
Bandwidth Limit (typical) 
20MHz     
Lower Frequency 
Response (AC –3dB) 
≤5Hz (at input BNC) 
Rise Time at BNC (typical) 
<3.5ns, <7ns, respectively at 100MHz, 50MHz     
Dynamic range 
± 5div 
DC Gain Accuracy 
2mV/div-5mV/div: 
± 4% (In Normal or Average acquisition mode) 
10mV/div-10V/div: 
± 3% (In Normal or Average acquisition mode) 
DC Measurement 
Accuracy, Average   
Acquisition Mode 
When vertical displacement is zero, and N ≥16: 
± (DC Gain Accuracy× reading+0.1div+1mV)   
When vertical displacement is not zero, and N ≥16: 
± [DC  Gain  Accuracy× (reading+vertical  displacement)+(1%  of 
vertical displacement) + 0.2div] 
When vertical scale is between 2mV/div and 245mV/div, add 2mV 
more for setting value.   
When vertical scale is between 250mV/div and 10V/div, add 50mV 
more for setting value.   
Delta Volts   
Measurement Accuracy   
(Average Acquisition Mode) 
Under  same  setting  and  condition,  the  voltage  difference  (△ V) 
between any two points in the waves coming from the average of 
more  than  16  waves  have  been  acquired:  ± (DC  Gain 
Accuracy× reading + 0.05 div) 
Trigger 
Trigger Sensitivity 
0.1div~1.0div (adjustable) 
Trigger Level Range 
Internal   
± 6 divisions from center of screen 
EXT   
± 1.2V 
Trigger Level Accuracy (typical) 
applicable  for  the  signal  of 
rising and falling time ≥20ns 
Internal 
± (0.3div ×  V/div)(± 4 divisions from center of screen) 
EXT 
± (6% of setting + 200 mV) 
Trigger Offset 
In Normal  mode: pre-trigger (memory depth/ 2*Sample  rate), 
delayed trigger 1s       
In Slow Scan mode: pre-trigger 6div, delayed trigger 6div   
Trigger Holdoff Range 
500ns~1.5s 
Set Level to 50% (typical) 
When input signal frequency ≥50Hz 
Edge Trigger   
Edge trigger slope 
Rising, Falling, Rising + Falling   
Pulse WidthTrigger   