Use Instructions
RIMS-IFU-001
/Rev.1 Page29/43
3. InstructiontoUse
① PreparethesterilizedEPIAdevice,anepiduralneedle,anda5mlsyringe.
Ifneeded,fillthesyringewithsalineorananesthetictobeinjectedintotheepiduralspace.
② Removeastyletfromtheepiduralneedle.
③ Combinethesyringeandtheepiduralneedle.
④ OpenthesyringecoverofEPIAbypressingtheOPEN button.
⑤ Insert the flange of thesyringebarrelintothesyringeholderofEPIAandplacetheneedleinthe
grooveinthefrontofthedevice.
WhenusingaTuohytypeneedle,inserttheneedleandsyringeinthecorrectdirectionconsidering
thebevel(slope)oftheneedletipandthecurveddirection.
Typically,thebevelfacestoanterior(cranial,towardhead)sothattheinjectedanestheticcaneasily
spreadintoanteriorepiduralspace.
⑥ ClosetheEPIAcoveruntilitclicksandsecurethesyringetoavoidshaking.
⑦ AfterconnectingEPIAtoadisplaydevice(Pleasereferto“section2.DeviceConnectionofChapter
III”),runtheprogramonthedisplaydevice.
⑧ Beforethestart,checkthatthegraphislocatedatthestart(setting)position.
(TheXandYaxesofthegraphintheprogramshouldbelocatedat0.)
AlsocheckthatthesyringeholderofEPIAislocatedatthestart(setting)position.
For the proper insertion, it is recommended to protrude a part of needle (about 20 mm) by
pressingtheForwardbuttonandthenStopbutton.
Usetheprotrudingpartoftheneedle,toinsertitintotheinjectionsiteinadvance.
④⑤⑥