Use Instructions

RIMS-IFU-001
/Rev.1 Page30/43

Inserttheprotrudingpartoftheneedleintothepatientsinjectionsite,andholdtheEPIAfirmly
againsttheoperationsite.
Hold the EPIA firmly and fix it onto the patient’s injection areawith one hand (lefthand
recommended).
OperatethebuttonsofEPIAwiththeotherhand(righthandrecommended).
BypressingtheForwardbutton, thesyringeandtheneedlewillautomaticallyadvance,and
theneedlewillbeinsertedintothepatient’sbody.
Duringtheoperation,makesurethatthedeviceispressedfirmlyagainsttheinjection
areatoavoidbeingpushedback.
Duringtheoperation,theplungerisnotaffected,therebytheanestheticisnotinjected
whiletheneedleisbeinginserted.
Thepressuredataaretransmittedinreal‐timetothedisplaydevice.
Throughthepressuregraph,theoperatorcanmonitortheintrainsertionmovementoftheneedle,
needleinsertionlength,andthepressurechangeappliedtotheneedle.
Whenthetipoftheneedleenterstheepiduralspace,thepressuredecreasesrapidly.
Whilemonitoringthechangesinthepressuregraph,theoperatorcanstoptheprogressoftheneedle
bypressingtheSTOPbutton .
Bydetectingarapiddropinpressure,thedevicewillautomaticallystopattheepiduralspace.
Theoperatorcanconfirmthatthe pressureisconstantlykeptlowbypressingthe0.2mm
button,therebyconfirmingthattheautomaticstoppointcoincideswiththeepiduralspace.
After the stop,if additional progress is required, press the 0.2mm button to advance the
needlefinely.
Ifawithdrawalisrequired,presstheBackbutton tomovetheneedlebackward.
