4. PrincipleofOperation
Theproductisusedtoassistwithanepiduralprocedurethatdetectstheepiduralspacebyautomatically
pushingthesyringebarreltoinserttheepiduralneedleintotheepiduralspace.
Themotorrotatesbyusingtheelectricpowerofa3Vbattery,andtherotationalmotionisconvertedtoa
linearmotionofthesyringebarrelandtheneedlefixedtothesyringeholder.
Whiletheepiduralneedleisbeinginsertedintothebody,thepressuresensorinthesyringeholderdetects
thepressurechangebetweeneachtissue.Thedetectedpressurechangeisconvertedandstoredasdigital
dataanddisplayedasagraphonthedisplaydevice.
When the epidural needle is inserted into the body, the pressure will gradually increase in the
subcutaneous()andligamenttissue(),andthepressurewillrapidlydropwhenitreachestheepidural
space().Inthisway,theoperatorscanverifywhethertheneedletiphassuccessfullyenteredintothe
epiduralspace.
EPIAalsomakesuseofthisphenomenonanddetectsreactionforce,whichismadeagainsttheforcethat
theneedlepenetratesthetissue.Whilethetipofneedleinsertsthroughthesubcutaneousandligament
tissue,thereactionforceaswellaspressureatthetipoftheneedleincreases(and inthegraph
below).When the tip of needle reaches the epidural space,the pressureand the reactionforce rapidly
decrease().ThepressuresensorofEPIAdetectsthischangeinreactionforceandconvertitasvisualdata.


Decrease of
∆𝒇
∆𝒕
at
indicates epidural space
auto-stop