Instruction manual

69
Measurement
When using the unit in a mode other than “sound level measurement”, all
processing functions provided by the unit (L
eq
, L
E
, L
max
, L
min
, L
N
) are car-
ried out simultaneously. (However, for the sub channel, only the additional
processing function set to “ON” in the menu list screen is carried out.) For
example, when equivalent continuous sound level measurement is selected,
the sound exposure level and percentile sound level are also determined.
However, the time percentage setting for the percentile sound level (5 values)
must be selected beforehand.
To display measurement values other than sound level, you need to set desired
items to ON on the [Display] screen from the menu list screen.
Also, make sure that the date and time are set correctly, as described on
page 25.
Sound level (L
p
) measurement
The procedure for sound level measurement is described below.
Preparations as described in the “Preparations” chapter must be completed
rst.
1. Press the POWER key to turn the unit on.
After the power-on screen has been shown, the measurement screen
appears.
The measurement parameter settings that were active before the unit
was turned off will be established again. Therefore the actual display
may not always be the same.