Instruction manual
38
Setting Measurement Parameters
TRIGGER MODE: OFF LEVEL EXTRN TIME
Sets the trigger mode for starting the measurement. (See pages
68 to 73.)
OFF: Trigger mode is disabled.
LEVEL: Trigger level can be set from 10 to 140 dB in 1 dB steps.
SLOPE: - or + are used to set the slope for level triggering.
EXTRN: Selects a CMOS logic-level signal to be used as trigger, sup-
plied to the trigger input on the side of the unit.
TRG TIME: Can be set from 00 h 00 m to 23 h 59 m in 1 minute steps.
PERIOD: Sets the trigger repeat interval. Can be set from 0 to 24 hours in
1 hour steps.
DELAY TIME: 0 to 10 S
Sets the delay between the point when the START/STOP key
is pressed and the actual start of measurement. Can be set from
0 to 10 seconds in 1 second steps. (See page 74.)
SUB FRQ. WEIGHT:FLAT A C
Sets the frequency weighting characteristic for the sub channel.
SUB TIME CONST: FAST SLOW 35mS 10mS IMPULSE
Sets the time weighting for the sub channel.
Condition Memory: 0: DEFAULT 1: P_ON 2-5: USER
Determines the way measurement parameter settings are stored
in memory.
0: DEFAULT Unit is reset to default measurement parameters when turned
off (except for SAVE, ERASE).
1: P_ON Measurement parameters are stored when unit is turned off and
reestablished when next turned on (except for SAVE,
ERASE).
2 to 5: USER User-selectable measurement parameter settings. Use the
LEVEL UP/DOWN keys to select settings.
Operation: LOAD SAVE ERASE
Controls the measurement parameter setting memory opera-
tion.
LOAD: Load the specified measurement parameter settings. Previ-
ously established settings are lost.
SAVE: (USER memory only)
Save the specified measurement parameter settings. The date
and time of the save operation are also stored.
ERASE: (USER memory only)
Serves to erase specified measurement parameter settings.