Instruction manual
73
Trigger Function
Time trigger
TIME TRG (Measurement start time): 00 h 00 m to 23 h 59 m in 1 minute steps
PERIOD (Measurement repeat interval): 0 h to 24 h in 1 hour steps
1. Set the measurement start time with the "TIME TRG" item.
2. Set the measurement repeat interval with the "PERIOD" item.
z When PERIOD is set to "0 H", the measurement is performed only once at the
preset time.
z When PERIOD is set to a value smaller or equal to the measurement time, trigger
events within the measurement time are disregarded.
An example for the time trigger operation is shown below.
Measurement start time: 1 h 30 m
Measurement is performed every 2 hours after that.