Operating instructions

Table Of Contents
Commissioning RIA452
68 Endress+Hauser
6.3.4 Digital input - DIGITAL INP./M5
The settings for the digital status inputs, e.g. for monitoring pumps, starting/stopping the counter
or resetting the min/max-value memory are grouped in this section.
Note: The digital status inputs are permanently assigned to the relays in the PUMP function. Relay 1
is monitored by digital input 1, relay 2 by digital input 2 etc.
Fail mode Hold
const
Min
Max
Output value if a sensor or device error occurs.
Hold = last valid value
Const = freely selectable value
Min = output value is 3.5 mA for 4-20 mA, and 0 V or 0 mA otherwise
Max = output value is 22.0 mA for 0/4-20 mA, and 1.1 V or 11 V otherwise
Fail value 0..999.99 The freely selectable value for "Fail mode = Const" can be set here.
Current output: 0...22 mA
Voltage output: 0...11 V
Simu mA OFF
0.0 mA
3.6 mA
4 mA
10 mA
12 mA
20 mA
21 mA
Outputs the selected current at the output regardless of the input value.
Is automatically set to OFF when exited.
Simu V OFF
0.0 V
5.0 V
10.0 V
Outputs the selected voltage at the output, regardless of the input value.
Is automatically set to OFF when exited.
Function (menu item) Parameter setting Description
Function (menu item) Parameter setting Description
Function OFF
Pump
Res. Tot.
Start/Stop
Min/Max
Function of the selected digital input.
•OFF
Pump = pump monitoring (see Pump monitoring function)
Res. Tot. = reset the totalizer*
Start/Stop = start or stop the totalizer*
Min/Max = reset the min/max memory values
!
Note!
Parameters marked * are only available with the pulse output option.
Level Low
High
Selects the side for evaluation.
Low = descending side
High = increasing side
Sampl. time 0..99 Defines the time within which pump feedback at the digital input is to be expec-
ted. If there is no feedback within the defined time, an error message is generated
and a second pump is activated if more than one pump is available.